服務熱線
400-617-8668
型號:TMDP0200
產品時間:2023-11-10
簡要描述:
TMDP0200高壓差分探頭用于測量兩個均非為地的測試點之間的電壓差。 泰克的高壓差分探頭可用于高達 6000 V 的信號。這類探頭由于具有共模抑制能力,成為較大部件中進行非地參考、浮動或隔離測量的JIA選擇。這些產品由泰克設計、制造和維修。
TMDP0200高壓差分探頭Tektronix offers a broad portfolio of high-voltage probing solutions that enable users to safely and accurately make floating measurements.
Key performance specifications
Key features
Applications
Connectivity
Safe high-voltage probe solutions
The THDP0100 and P5210A have the largest differential dynamic range capability from Tektronix, allowing users to safely measure up to ±6000 V. The THDP0100 supports bandwidths up to 100 MHz and slew rates up to 2500 V/ns at 1/1000 gain. These probes are supplied with two sizes of hook tips and have an overrange visual and audible indicator which warns the user when they are exceeding the linear range of the probe.
The THDP0200 and P5205A are active differential probes that are capable of safely measuring differential voltages up to ±1500 V. The probes are effective in making measurements in IGBT circuits such as motor drives and power converters. The THDP0200 supports bandwidths up to 200 MHz and slew rates up to 275 V/ns at 1/250 gain.
TMDP0200高壓差分探頭 are designed for medium-voltage applications with differential requirements up to ±750 V. These probes have lower attenuation ranges and offer better signal-to-noise ratio.
The P5200A can be used with any oscilloscope and enables users to safely make measurements of floating circuits with their oscilloscope grounded. The P5200A Active Differential Probe converts floating signals to low-voltage ground-referenced signals that can be displayed safely and easily on any ground-referenced oscilloscope.
Warning: For safe operation, do not use the P5200A High-voltage Differential Probe with oscilloscopes that have floating inputs (isolated inputs), such as the Tektronix TPS2000 and THS3000 Series oscilloscopes. The P5200A High-voltage Differential Probe requires an oscilloscope or other measurement instrument with grounded inputs.
Connectivity options
The TMDP and THDP Series probes are equipped with the TekVPI®interface which allows smart communication between the oscilloscope and probe. Pushing the probe menu button will launch the probe control menu on the oscilloscope display providing access to all relevant probe settings and controls. These probes are designed to operate on TekVPI®oscilloscopes without requiring the TPA-BNC adapter.
The P52xxA Series probes are equipped with the Tektronix TekProbe®interface which communicates scale information to the oscilloscope. Direct connections can be made to oscilloscopes configured with the TekProbe®interface or to any oscilloscope when used with the 1103 TekProbe®Power Supply.